Error Code |
Meaning |
Suggested Repair |
FW Version of Error |
E001 |
Serial number is missing or corrupt |
Return to LRS for repair |
1.26 |
E002 |
Pager type is invalid |
Return to LRS for repair |
1.26 |
E003 |
Radio calibration is missing or corrupt |
Return to LRS for repair |
1.26 |
E004 |
Frequency not supported by firmware |
Return to LRS for repair |
1.26 |
E005 |
Frequency not supported by hardware |
Return to LRS for repair |
1.26 |
E006 |
Calibration does not support frequency |
Return to LRS for repair |
1.26 |
E007 |
System ID is invalid |
Perform OTA pager programming |
1.26 |
E008 |
Pager ID is invalid |
Perform OTA pager programming |
1.26 |
E009 |
Staff pager mode not supported |
Perform OTA pager programming and program as guest pager |
1.26 |
E010 |
System ID is locked |
The system ID used to program the pager does not match that of the pager. Attempt to program using the right system ID or return to LRS for repair. |
1.37 |
I see Error Code E001?
The pager serial number is invalid or corrupted.
Requires pager sent back to LRS.
I see Error Code E002?
The pager type is incorrect (ie: CS7 is told to behave as a CS6).
Requires pager sent back to LRS.
I see Error Code E003?
The pager’s radio calibration/tuning is missing or corrupt.
Requires pager sent back to LRS.
I see Error Code E004?
The selected frequency is invalid.
Requires pager sent back to LRS.
I see Error Code E005?
The selected frequency is invalid. Requires pager sent back to LRS.
I see Error Code E006?
The calibration will not work with the selected frequency.
Requires pager sent back to LRS.
I see Error Code E007?
The pager was programmed with an invalid System ID.
User can reprogram the pager in the field with LRS Transmitter. Follow transmitter instructions for “reprogramming a guest pager”.
I see Error Code E008?
The pager was programmed with an invalid Pager ID.
User can reprogram the pager in the field with LRS Transmitter. Follow transmitter instructions for “reprogramming a guest pager”.
I see Error Code E009?
The pager was programmed with as a Staff Pager (ex: SP4).
User can reprogram the pager in the field with LRS Transmitter. Follow transmitter instructions for “reprogramming a guest pager”.
I see Error Code E010?
The pager was configured to only accept programming commands for one System ID. The attempted programming failed because it was for a different System ID. The error will disappear at the end of the reprogramming window of 30 seconds and return to displaying it's number.
If paged while showing the error, the pager number shows back up immediately. The error will also disappear if placed on charge.
User can verify the system ID used to program the pager.